UConn Health Center HomeFlow Cytometry
HOME OTHER HEALTH CENTER CORES DEPARTMENT OF IMMUNOLOGY CONTACT US
 
 

LSR IIA Laser Configurations and Possible Fluorochrome Usage


Becton-Dickinson LSR IIA

Laser Line
Detector
Long Pass Filter
Band Pass
Fluor
488nm blue laser A 735 780/60 PE-Cy7, PE-Alexa 750
B 685 695/30 PerCP-Cy5.5, PE-Cy5.5, 7AAD, PE-Alexa 700, Alexa 680
C 635 670/14 PerCP, PE-Cy5, PE-Alexa 647, 7AAD, Fura-Low-Ca++
D 600 610/20 PI, PE-Texas Red, PE-Alexa 610, PE-Alexa 594
E 550 575/26 PE
F 505 530/30 Fitc, Alexa 488, YFP, GFP, CFSE, Calcein AM
G - 488/10 SSC
633nm red laser A 735 780/60 APC-H7, APC-Cy7, APC-Alexa 750
B 680 695/40 APC-Cy5.5; Alexa 700 with 710/20 filter
C - 660/20 APC, Alexa 647
405nm violet laser A 505 585/42 Pacific Orange, Alexa 430 with 525/50 filter
B - 440/40 Alexa 405, CFP, Pacific Blue
C - - -
325nm (LSR IIA); 355nm (LSR IIB) UV Laser A 505 525/50 Indo1 Blue
B - 440/40 DAPI, Indo1 Violet, Hoechst 33342
C - - -
  